While being transferred from a surgical table to a patient lift after her surgery, Ms. Feldman was dropped on the floor.  Simultaneously, the top of the surgical table fell to the floor.

We investigated this loss, tested the table operation, demonstrated the sequence of events with a crash-test dummy on the same equipment, and demonstrated that a loose table top could not have caused the fall.  The fall was caused by tightening one side of the patient lift sheet while the other side was loose; this caused the patient to roll off the side of the table.




photo1 photo2 photo3
photo 4 photo 5 video